888-24-314-G-00 rev. B • 12/14
Medical-Grade
Power Strip
97-466
Fully tested, certified and compliant to:
·
UL 1363A
Input and Output: 120 V~, 60 Hz, 15 A max
822-104-01
Made in China
yyww
ENGLISH
Maintenance & Safety
Earth Bond Test: To ensure safety grounding between the
power cord ground connection and any accessible metal
parts on the cart, although not mandatory, the following test
is recommended to be performed every two years using a
calibrated medical device safety analyzer. The procedure is as
follows:
1. Disconnect the cart from mains power (unplug cart from the
wall).
2. Remove power from internal power system outlets by
pressing the power button on the Power System Interface.
3. Set up Earth Bond Test per the medical device safety analyzer
instructions.
4. Connect the Cart power cord to the medical device safety
analyzer.
5. Remove cover to access Lift Tension Adjustment point.
6. Attach second medical safety device analyzer probe to Lift
Tension Adjustment point on Cart.
7. Perform Test (25 amps).
8. Ensure Earth Bond is less than or equal to 0.2-ohms.
9. Remove medical device safety analyzer connections to Cart,
replace cover over Lift Tension Adjustment point, return the
Cart to service.
Insulation Resistance Test: To ensure mains power lines are
adequately insulated from earth ground, the following test,
although not mandatory, is recommended to be performed
annually using a calibrated medical device safety analyzer. The
procedure is as follows:
1. Disconnect the cart from mains power (unplug cart from the
wall).
2. Remove power from internal power system outlets by
pressing the power button on the Power System Interface.
3. Set up the Insulation Resistance Test per the medical device
safety analyzer instructions.
4. Connect the Cart power cord to the medical device safety
analyzer.
5. Perform test (500 V).
6. Ensure Insulation Resistance is greater than or equal to 1
Mohms.
7. Remove medical device safety analyzer connections to Cart
and return Cart to service.
Recommended Periodic Inspection and Maintenance
• Inspect Cord Thoroughly Before Each Use. DO NOT USE IF DAMAGED.
• Do Not Plug More Than SPECIFIED NUMBER OF WATTS Into Power System.
• Do Not Run Cord Through Doorways, Holes in Ceilings, Walls or Floors.
• FULLY INSERT Plug Into Outlet.
• Do Not Remove, Bend or Modify Any Metal Prongs or Pins of Cord.
• Do Not Use Excessive Force to Make Connections.
• Keep Away From Water. DO NOT USE WHEN WET.
• Keep Children Away From Cord.
• Do Not Plug Cord into An Extension Cord.
• AVOID OVERHEATING. Uncoil Cord and Do Not Cover It With Any Material.
• Do Not Drive, Drag or Place Objects Over Cord. Do Not Walk On Cord.
• GRASP PLUG to REMOVE FROM Outlet. Do Not Unplug By Pulling On Cord.
Polarized Connector Inspect for damage. Weekly Any user.
Power Outlets Inspect for damage or dirt. Weekly Any user.
Circuit Breakers Test Circuit Breakers. Weekly IT Personnel
Coiled Cord
Inspect for wear, damage, or
stretching. Move attachment
hook to avoid dragging on  oor
Weekly Any user
Power Strip Internal & External Components Disposal
1. Internal wiring, power cord and all electrical components should be recycled through an electronics recycler.
2. Remaining plastics and metals can be recycled through a commercial recycler.
Recommended Periodic Inspection and Maintenance
Component Action How often By whom
UI, Ethernet,
USB cables
Inspect for wear,
pinching, bad connectors
Monthly Any user
Casters
Inspect for wear and
debris
Monthly Any user
Maximum
Load
Inspect to ensure that
maximum recommended
loads are not exceeded
Daily Any user
Power System
SLA batteries
Replace lead acid
batteries
1,2
It is recommended that a discharge
test be conducted when cart
is  rst purchased to determine
runtime with computer equipment
con gured on the cart. Change
batteries when instructed by
StyleLink battery management
software. If StyleLink is not used,
it is strongly recommended that
a monthly discharge test be
conducted to determine runtime
loss. If runtime is less than 50% of
original, batteries must be changed.
IT Personnel
Power Module
Inspect for dust at intake,
vacuum as required using
a vacuum cleaner that
DOES NOT generate ESD
(Electrostatic Discharge)
Monthly IT Personnel
Battery
harness
Inspect for wear,
cracking, pinching, or
other damage
Monthly IT Personnel
Coiled Cord
Inspect for wear, damage,
or stretching.
Weekly Any user
1. Please dispose of all batteries in accordance with local law
2. Always replace with similar size battery (33 A-h to replace 33 A-h) and always replace in pairs